150 Guidance and manufacturer's declaration
Recommended separation distances between portable and mobile RF communications
equipment and the monitor
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
0.36894
0.36894
1.1667
1.1667
3.6894
3.6894
11.667
11.667
®
Connex
0.73785
2.3333
7.3785
23.3333
Spot Monitor