Directions for use
Message
Tube type does not
match device
configuration.

SpO2 messages

General SpO2 messages
Message
SpO2 not functional.
044800
SpO2 rebooting. 044900 The SpO2 module is rebooting
SpO2 rebooting. 044901 The SpO2 module stopped
SpO2 rebooting. 044902 SpO2 received a packet with bad
SpO2 rebooting. 044903 The SpO2 power on self test
SpO2 rebooting. 044904 The SpO2 power on self test
Masimo messages
Message
Sensor not connected.
Clear to retry. 040600
Replace the SpO2 cable.
040700
Possible cause
The tube type settings and the
actual tube type do not match
Possible cause
The SpO2 power on self test
failed
sending data
CRC from module
failed
timed out
Possible cause
The SpO2 cable is not connected
The SpO2 cable life has expired
Suggested action
Change the tube type setting to
match the actual tube type.
Suggested action
Internal hardware malfunction in
SpO2 module. Replace the module.
Informational error. The host
software is attempting to clear an
error by rebooting the SpO2 module.
No action required.
Informational error. The host
software is attempting to clear an
error by rebooting the SpO2 module.
No action required.
Informational error. The host has
received a packet with bad CRC from
the SpO2 module. The packet in
question is ignored. No action
required.
Internal hardware malfunction in
SpO2 module. Replace the module.
Internal hardware malfunction in
SpO2 module. Replace the module.
Suggested action
Connect SpO2 cable. If the problem
is still present, replace the SpO2
sensor. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
Troubleshooting 95
Alarm priority
Information
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Alarm priority
Very low
Very low