90 Troubleshooting
Message
Unable to determine
NIBP; check
connections; limit
patient movement.
050003
Unable to determine
NIBP; check
connections; limit
patient movement.
050004
Unable to determine
NIBP; check inflation
settings. 050005
Unable to determine
NIBP; check connections
and tubing for kinks.
050006
Unable to determine
NIBP; check
connections; limit
patient movement.
050007
Unable to determine
NIBP; check
connections; limit
patient movement.
050008
Unable to determine
NIBP; check inflation
settings. 050009
Unable to determine
NIBP; check
connections; limit
patient movement.
05000A
Unable to determine
NIBP; check inflation
settings. 05000B
Unable to determine
NIBP; check connections
and tubing for kinks.
05000C
No display
Unable to determine
NIBP; check
connections; limit
Possible cause
The NIBP measurement may be
inaccurate, patient motion
occurred, or the settings for
patient readings obtain might not
be accurate
Excessive artifact, no blood
pressure parameters could be
calculated
Low inflation on blood pressure
measurement attempt
The NIBP tubing has a kink, or
there is a NIBP transducer
calibration failure
The blood pressure measurement
dumped too soon
There were not enough steps on
the measurement attempt
There is invalid patient
information for the selected mode
The re-inflation was too late on
the measurement attempt
There were many re-inflate
attempts on the measurement
attempt
Couldn't bleed pressure below
safe venous return pressure
The safety check failed on the
measurement attempt
The NIBP pressure is not stable
and the transducer zero value
cannot be set
Suggested action
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Unable to determine blood pressure.
Check connections; restrict patient
movement. If problem persists,
replace module
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Unable to determine blood pressure.
Check connections; restrict patient
movement.
Make sure the NIBP settings/patient
mode is appropriate. If the problem
persists, replace the module.
Unable to determine blood pressure.
Check connections; restrict patient
movement.
Unable to determine blood pressure.
Check connections; restrict patient
movement.
Unable to release cuff pressure.
Check tubing for kinks and
connection integrity.
Internal malfunction. If the problem
persists, replace the module.
The NIBP pressure is not stable and
the transducer zero value cannot be
set. If problem persists, replace
module.
®