58 Patient monitoring
Stat intervals
You can configure the monitor to take NIBP measurements continuously.
When you select Stat on the Intervals tab in Settings, the monitor takes repeated NIBP
measurements for 5 minutes, starting a new cycle each time the cuff deflates below
safe venous return pressure (SVRP) for 2 seconds.
Current cuff pressures are not dynamically displayed during a Stat reading. The Home
tab displays the NIBP reading from the previous cycle until the current cycle finishes.
Note
Start Stat intervals
1. Place the proper cuff around the patient's bare upper arm.
2. On the Home tab, touch
The Intervals screen on the Settings tab appears.
3. Select Stat.
4. Touch Start intervals.
Averaging intervals
The averaging interval program enables you to record the patient's average NIBP
readings over a set period of time.
Start Averaging intervals
Note
Note
1. Place the proper cuff around the patient's bare upper arm.
2. On the Home tab, touch
The vertical Intervals tab on the Settings tab appears.
3. Select the desired program. Touch the program you want to use. For example, touch
Program 2.
4. Touch Start intervals.
The Program name appears on the Home tab, along the averaged reading as the
readings occur.
5. Touch Save after the Averaging intervals completes.
WARNING Patient injury risk. If you use Stat mode repeatedly, periodically
observe the patient's limb to ensure that circulation is not impaired and that
the cuff remains in place. Prolonged impairment of circulation or improper
cuff position can cause bruising.
Touch STOP to stop intervals. To restart intervals, go back to the Stat
intervals screen.
.
You must be in the Office profile to access Averaging intervals.
Authorized personnel can configure the Averaging intervals in Advanced
settings.
.
®