26 Startup
Type of alert
Low Alarm
Very Low Alarm
Information message
Press
When the device is powering up, the LED flashes until the monitor displays the
Welch Allyn logo and a power-up tone sounds.
A pop-up screen appears, depending on your configuration and functionality.
On initial power-up, the monitor prompts you to set the language, date, and time.
See "Change the language" and "Set the date and time" for directions.
If your facility has chosen a login format, then the first image you see is the login
screen.
If your facility has not chosen a login format, then the first image you see is the
Home tab.
If Bluetooth
is available.
®
Bluetooth
wireless technology
Note
Bluetooth wireless technology is available in the Office profile.
WARNING Steady amber indicates a low-level alarm. Flashing amber
indicates a medium-level alarm. Flashing red indicates a high-level alarm.
WARNING Always observe the monitor during power-up. If any display
fails to illuminate properly, or if a system fault code or message displays,
inform qualified service personnel immediately, or call your nearest Welch
Allyn Customer Service or Technical Support facility. Do not use the
monitor until the problem is corrected.
CAUTION Always use the monitor with an adequately charged and
properly functioning battery.
CAUTION During intervals monitoring, keep the monitor connected to AC
power at all times.
CAUTION Use only a Class I (grounded) AC power cord to charge the
battery for this monitor.
to power up the monitor.
®
is enabled, a list of paired devices and the option to add a new device
Your model might not contain all of these features.
Color
Steady amber
Cyan