86 Maintenance and service
7. Power on the monitor and verify that the monitor functions normally before using it.
If liquids possibly entered the monitor, remove the monitor from use until it has been
properly dried, inspected, and tested by qualified service personnel.
Clean the equipment
The screen lock blocks the display of patient information and prevents any input, which
may be useful when cleaning the display.
Follow the cleaning agent manufacturer's instructions to prepare solution, if applicable,
and clean all exposed surfaces of the monitor, APM work surface, accessory bin(s) and
basket, cords and cables, and stand. Wipe all surfaces until no visible soil remains.
Change the wipe or cloth throughout the cleaning procedure as needed.
1. Disconnect the AC power cord from the mains outlet.
2. Wipe the top of the monitor.
3. Wipe the sides, front, and rear of the monitor.
4. Avoid residual film buildup on the LCD screen. After cleaning and disinfecting, wipe
the LCD screen with a clean cloth dampened with water, and then wipe the screen
dry with a dry clean cloth.
5. Wipe the bottom of monitor.
6. Wipe the APM work surface.
7. Wipe the accessory bins or basket.
8. Wipe the AC power cord and the APM work surface power/USB cable assembly.
9. Wipe the stand from top to bottom.
Dry the equipment
1. Allow all components except the LCD screen to air dry.
2. Wipe the LCD screen dry with a clean cloth.
Store the device
Store the device according to facility guidelines to keep the device clean, dry, and
ready for service.
Cleaning accessories
Accessories include components like blood pressure cuffs and hoses, SpO2 sensors and
cables, thermometers, and the barcode scanner. Follow accessory manufacturer's
instructions for cleaning and disinfection.
WARNING Electric shock hazard. Do not open the monitor or attempt
repairs. The monitor has no user-serviceable internal parts. Only perform
routine cleaning and maintenance procedures specifically described in this
manual. Inspection and servicing of internal parts shall only be performed
by qualified service personnel.
CAUTION Sterlizing the monitor could damage the device.
®