122 Troubleshooting
Message
Certificate load
successful.
Certificate load failed.

Connectivity messages

Message
Unable to obtain wired
device IP address.
210000
Network not found;
check network cable
connection. 210100
Unable to communicate
with NRS. 360000
Communication error
with host. 1A0000
Data rejected. CRC
mismatch. 1A0001
Data rejected.
Unsupported message.
1A0002
Data rejected. Invalid
parameter. 1A0003
Data rejected.
Deserialize the object.
1A0004
Data rejected.
Unsupported message.
1A0005
Data rejected. Patient ID
required. 1A0006
Data rejected. Clinician
ID required. 1A0007
Data rejected. Time
mismatch. 1A0008
Possible cause
The radio customer certificate was
successfully loaded from the host
The radio customer certificate was
not loaded
Possible cause
No wired connection
Lost ethernet DHCP address
Cannot communicate with NRS
Communication timeout failure
with external Host
The WACP stack has detected a
CRC mismatch in the message
External Host NACK - The host
does not support the message/
object.
The message has an invalid
parameter.
The monitor failed to deserialize
the object.
The host is in a state that cannot
accept the message.
The message has a missing
patient ID
The message is missing a clinician
ID
The message has a mismatched
time
Suggested action
None.
Try again.
Suggested action
Verify DHCP functionality and configuration.
Check wired connection into device then
verifiy DHCP functionality and configuration.
Verfiy NRS IP configuration and functionality. Very low
Verify the external host services are loaded
and started on the server. If the problem
persists, check for available software
upgrades of the monitor or the system.
Check the data and try again. If the problem
persists, contact your system administrator.
Check the monitor and try again. If the
problem persists, contact your system
administrator.
Check the data and try again. If the problem
persists, contact your system administrator.
Check the data and try again. If the problem
persists, contact your system administrator.
Check the data and try again. If the problem
persists, contact your system administrator.
Add the patient ID to the record.
Add the clinician ID to the record.
Ensure that the clock on the monitor and the
server match.
®
Connex
Spot Monitor
Alarm
priority
Information
Information
Alarm
priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low