Service Manual
Opening the SpO
1.
Separate the SpO
Module"
2. Two tabs secure the SpO
small screwdriver or your finger, press on each of the tabs inside the SpO
chassis while applying pressure to separate the rear panel from the chassis, as shown
here. (When separating the rear panel from the chassis, make sure the side panels
remain seated in the rear panel.)
SpO2 Module
Rear Panel
3. Once the tabs allow the rear panel to be removed, separate the rear panel from the
chassis.
The rear panel is sealed to the chassis with rubber O-ring gaskets. These gaskets can
be reused when re-assembling the SpO
SpO2
Connector
Panel
When closing the SpO
becomes pinched between the rear panel and the chassis.
Module
2
module from the monitor
2
on page 55).
module rear panel to the SpO
2
Release
Continuous
Tabs
O-Ring Gasket
SpO2
Board
Rear
Panel
O-Ring
Cord
Release
Tab (2)
module, verify no O-ring gasket slips out of its channel and
2
("Removing the Dual Battery in SpO2
module chassis. Using a
2
module.
2
Chassis
SCP Board
Speaker
0696-53
Repair Procedures