4
Probe Calibration
To Calibrate the Probe on InfiniiVision Oscilloscopes
To Calibrate the Probe on InfiniiVision Oscilloscopes
Calibration Procedure
70
The following test equipment are needed for the calibration
procedure.
Table 9
Required Calibration Test Equipment
Test Equipment
N2825A User-Defined
Probe Head
Pomona 3781-12-2 Red
Minigrabber Test Clip
Patch Cord
Pomona 3781-12-0 Black
Minigrabber Test Clip
Patch Cord
1. Connect the probe to the oscilloscope
a Turn the oscilloscope on.
b Plug the primary channel of the N2820A probe into the
oscilloscope channel you are intending to use.
c If applicable, plug the secondary channel of the probe into the
oscilloscope channel you are intending to use.
d Configure the zoomed-in path to the primary or secondary
channel as necessary.
e Solder the sensor leads onto the N2825A user-defined head as
explained in
f Attach the N2825A user-defined head to the probe.
Description
Provided with probe.
"Quick Start"
on page 23.
Refer to
http://www.pomonaelectronics.com/
more information and how to order this
product.
Refer to
http://www.pomonaelectronics.com/
more information and how to order this
product.
N2820/1A User's Guide
for
for