Zoomed-In Input Current Range Versus Supply Voltage

N2820/1A User's Guide
Zoomed-In Input Current Range Versus Supply Voltage
For the zoomed-in probe input, the following two graphs show the
typical input current range allowed based on the supply voltage
that you are measuring. For example, when using an N2824A head
and measuring a 5V input, the current range is ∴0.54A. When
using an N2822A head and measuring a 5V input, the current
range is ∴1.0A.
Figure 38
Zoomed-In Input Current Range Versus Supply Voltage
(N2824A 100 mΩ Head)
Characteristics and Specifications