N2820A Probe
Black Test Clip of
Pomona 3788
N2820/1A User's Guide
4
Connect the equipment as shown in
the following requirements:
• Connect the probe to the scope channel that you will use during
probing. Use any other available scope channel for the BNC
cable that is needed for connection to the Pomona 3788.
• Connect the Pomona 3788 to two BNC cables using a BNC tee.
Then connect one of these BNC cables to a scope channel and
the other BNC cable to the Aux Output BNC connector located
on the Oscilloscope's rear-panel (as shown in
• On N2820A probes, do not connect the secondary cable at this
time.
• Connect both the N2825A's sensor leads (red and black wires) to
the black test clip of the Pomona 3788 as shown in
N2825A Probe Head
Figure 30
Example of Initial Calibration Setup
To Calibrate the Probe on Infiniium Oscilloscopes
Figure 30
Red Test Clip of
Pomona 3788
Probe Calibration
on page 67. Observe
Figure
30).
Figure
BNC
Cable
BNC Tee
BNC
Cable
Pomona 3788
4
30.