Average Gain of Zoomed-Out Channel

N2820/1A User's Guide
Perform this procedure only for N2820A probes.
Allow the probe to warm up for at least 20 minutes.
NOTE
1
Snap an N2825A head onto the N2820/1A probe and connect the
test equipment as shown in
positive side of the DC supply to the red lead on the N2825A head.
Figure 51
2
Set the DC power supply to +500 mV.
Although this is the recommended setting, you can use any voltage up to
NOTE
+700 mV.
Average Gain of Zoomed-Out Channel
Figure
Test Setup for Measuring V
Performance Verification
51. Be sure to connect the
in
7
101