Keysight N2804/5A Differential Probes User’s Guide Introduction The N2804A and N2805A differential probes provide superior general-purpose differential signal measurements that are required for high-speed power measurements, vehicle bus measurements, and high-speed digital system designs. Both probes are powered by the oscilloscope’s AutoProbe interface. Both probes are designed for Measurement Category II (CAT II) as described in “IEC...
Sales Office. If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Keysight Technologies Sales Office. Keep the shipping materials for the carrier’s inspection. The Keysight Technologies office will arrange for repair or replacement at Keysight Technologies’ option without waiting for claim settlement.
Observe all of the product limits noted on the products labels and listed in “Safety CAUTION Information" on page 17 and “Specifications and Characteristics" on page 20. Figure 2 N2805A Product Labels Figure 3 N2804A Product Labels N2804/5A Differential Probes User’s Guide...
N OTE support your probe, fix known defects, and incorporate product enhancements. To download the latest firmware, go to www.keysight.com, locate the Technical Support page, and search by your oscilloscope model number. The N2804A and N2805A probes are designed for oscilloscopes with 50W N OTE AutoProbe-interface channel inputs.
Disconnect the probe from the oscilloscope and clean the probe with a soft cloth dampened with a mild soap and water solution. Make sure that the probe is completely dry before reconnecting it to an oscilloscope. N2804/5A Differential Probes User’s Guide...
35. The measurement category of a combination of a PROBE ASSEMBLY and an accessory is WARNING the lower of the measurement categories of the PROBE ASSEMBLY and of the accessory. Figure 4 N2804A in Case N2804/5A Differential Probes User’s Guide...
Page 11
Pincer Clips (1 red and 1 black) Rating: 300V, 1A (CAT II) BW: 300 MHz Extension Leads (1 red and 1 black) Approximately 30 cm length BW: 120 MHz Probe Offset Adjustment Tool (Qty. 1) N2804/5A Differential Probes User’s Guide...
35. The measurement category of a combination of a PROBE ASSEMBLY and an accessory is WARNING the lower of the measurement categories of the PROBE ASSEMBLY and of the accessory. Figure 5 N2805A in Case N2804/5A Differential Probes User’s Guide...
This probe is not for electrically insulating the circuit under test and the measuring instrument. N2804A N2804A 300 MHz Differential Probe N2805A Terminate with 50 Ω N2805A 200 MHz Differential Probe Terminate with 50 Ω Figure 6 Attaching an Accessory Tip N2804/5A Differential Probes User’s Guide...
Using the probe’s hook tips, short the + and - probe inputs together as shown in Figure Figure 7 Adjusting the Offset Using the Supplied Tool Press Default setup and AutoScale of the oscilloscope. Press the channel button for the probe, and set the oscilloscope channel to DC coupled mode. N2804/5A Differential Probes User’s Guide...
Page 16
Set the vertical scale of the oscilloscope to 100 mV/div (for N2804A) and 50 mV/div (for N2805A). Using the offset adjustment tool that comes with the probe, adjust the probe offset voltage to zero volts. N2804/5A Differential Probes User’s Guide...
PROBE ASSEMBLY and of the accessory. Do Not Operate Without Covers. To avoid electrical shock or fire hazard, do not operate WARNING this probe with the covers removed. N2804/5A Differential Probes User’s Guide...
Page 18
The probe cable is a sensitive part of the probe and, therefore, you should be careful not CAUTION to damage it through excessive bending or pulling. Avoid any mechanical shocks to this product in order to guarantee accurate performance and protection. N2804/5A Differential Probes User’s Guide...
Page 19
Periodically inspect the probe wires to check for any cracks, damage, or defective WARNING leads. Do Not Operate With Visible or Suspected Damage. If you suspect there is damage, have it inspected by a Keysight authorized service personnel. Capacitors inside the instrument may retain a charge even if the instrument is WARNING disconnected from its source of supply.
Ensure that the environmental conditions do not exceed the probe’s specified limits. Table 5 N2804/5A Safety and Regulatory Information Description CEI/IEC 61010-031 CAT II This symbol indicates the Environmental Protection Use Period (EPUP) for the product’s toxic substances for the China RoHS requirements.
Page 21
–50 dB at 10 MHz –50 dB at 10 MHz Noise Referenced to Probe Input 90 mV 30 mV Power Requirements AutoProbe Interface AutoProbe Interface * All entries are typical unless otherwise noted. † Warranted specification ‡ Calculated 0.35/Tr N2804/5A Differential Probes User’s Guide...
Page 22
350 mm x 270 mm x 80 mm 410 mm x 365 mm x 65 mm (L x W x H) (13.8 in x 10.6 in x 3.15 in) (16.1 in x 14.4 in x 2.56 in) N2804/5A Differential Probes User’s Guide...
N2804A Performance Test Record Test Test Limits Test Results DC Differential Gain ±1% (99 mV – scope vertical accuracy to 101 mV + scope vertical accuracy) Accuracy ¡ Bandwidth (210 mV – scope vertical accuracy) N2804/5A Differential Probes User’s Guide...
Measurement Category IV (CAT IV) Measurement category IV is for measurements performed at the source of the low-voltage installation. For example, electricity meters and measurements on primary over-current protection devices and ripple control units. N2804/5A Differential Probes User’s Guide...
If the probe is found to be defective we recommend sending it to an authorized service center for all repair and calibration needs. Perform the following steps before shipping the probe back to Keysight Technologies for service. Contact your nearest Agilent sales office for information on obtaining an RMA number and return address.
Returning the Probe for Service Contacting Keysight Technologies For technical assistance, contact your local Keysight Call Center. • In the Americas, call 1 (800) 829-4444 • In other regions, visit http://www.keysight.com/find/assist • Before returning an instrument for service, you must first call the Call Center at 1 (800) 829-4444.
Page 38
Returning the Probe for Service N2804/5A Differential Probes User’s Guide...