CAUTION
CAUTION
N2820/1A User's Guide
If you are not skilled in soldering, ask a trained assembler to change the
resistor.
Figure 29
Location of R
5
Replace the head's cover using the four screws. (M1.6 x 0.35, 5 mm
long, Keysight part number 0515-5210)
To avoid damaging the aluminum casting, do not over tighten the screws.
6
Connect the N2825A to an N2820/1A probe and connect the probe
to the oscilloscope.
7
On Infiniium oscilloscopes,
a Click the Setup > Probe Configuration menu command.
b Select the tab for the probe's primary channel. N2821A probes
only have a primary channel.
c In the User Def Head Resistance field, enter the resistance of the
DUT's R
SENSE
d In the dialog box, click Add Head to give your head with
associated external R
To Add a Custom Internal R
Resistor (R1)
SENSE
resistor in ohms.
resistor a unique name.
SENSE