Filter, Averaging Type (Avt); Filter, Averaging Number (Avn) - MICRO-EPSILON capaNCDT 6200 Operating Instructions Manual

Hide thumbs Also See for capaNCDT 6200:
Table of Contents

Advertisement

Ethernet Interface
6.4.4
Mode of measurement averaging
Command
Response
Index
Request averaging type
Command
Response
6.4.5
Number of measuring values used to calculate the average (adjustable from 2 ... 8)
Command
Response
Index
Request averaging number
Command
Response
capaNCDT 6200

Filter, Averaging Type (AVT)

AVT
$AVTn<CR>
$AVTnOK<CRLF>
n = 0: No averaging (Default setting)
n = 1: Moving average
n = 2: Arithmetic average (outputs only each nth measuring value)
n = 3: Median
n = 4: Dynamic noise rejection
? = Request averaging type
$AVT?<CR>
$AVT?nOK<CRLF>

Filter, Averaging Number (AVN)

AVN
$AVNn<CR>
$AVNnOK<CRLF>
n = 2 ... 8
? = Request averaging number
$AVN?<CR>
$AVN?nOK<CRLF>