Functional Principle, Technical Data
2.3
Technical Data
Controller type
Resolution static
Resolution dynamic
Bandwidth
Bandwidth adjustable
Data rate output digital
Linearity (typical)
Max. sensitivity deviation
Long term stability
Synchronous operation supported
(multiple control units)
Isolator measurement
Temperature stability
Operating temperature, sensor
Operating temperature, controller
Storage temperature
Supply
Power consumption at
24 VDC
Output
capaNCDT 6200
DT62x0 with DL6220
0.02 % FSO (5 kHz)
max. 3.906 kSa/s
≤ ±0,05 % d.M.
≤ ±0.1 % FSO
≤ 0.02 % FSO / month
-50 ... +200 °C (-58 ... +392 °F)
+10 ... +60 °C (+50 ... +140 °F)
-10 ... +75 °C (+14 ... +167 °F)
DT6220: 24 VDC (12 ... 36 VDC)
DT6230: 24 VDC (15 ... 36 V)
DT6220
DT6230
per DL6220
per DL6230
0 ... 10 V (short-circuit proof)
4 ... 20 mA (load max. 500 Ohm)
0.004 % FSO
5 kHz (-3dB)
to 20 Hz
no
no
200 ppm
3.1 W (typical)
3.8 W (typical)
1.8 W (typical); 2.0 W (max.)
1.9 W (typical); 2.2 W (max.)
Ethernet
DT62x0 with DL6230
0.0005 % FSO
0.005 % FSO (5 kHz)
5 kHz (-3dB)
to 20 Hz
max. 3.906 kSa/s
≤ ±0,025 % d.M.
≤ ±0.1 % FSO
≤ 0.02 % FSO / month
no
no
200 ppm
-50 ... +200 °C (-58 ... +392 °F)
+10 ... +60 °C (+50 ... +140 °F)
-10 ... +75 °C (+14 ... +167 °F)
24 VDC (15 ... 36 VDC)
0 ... 10 V (short-circuit proof)
4 ... 20 mA (load max. 500 Ohm)
Ethernet