N2818A Performance Verification - Keysight N2818 User Manual

Differential probes
Table of Contents

Advertisement

Using the Probe

N2818A Performance Verification

NOTE
Description
Digitizing Oscilloscope
Signal Generator
Precision DC voltage source
BNC Adapter
50Ω BNC Feed Through Adapter
36
The following procedure can be used to test the N2818A probe's DC
differential gain accuracy and bandwidth, which are warranted
specifications.
The recommended test interval is 1 year.
Table 11
Required Test Equipment
Critical Specifications
Bandwidth: >200 MHz
1 MΩ/50Ω selectable input
Amplitude accuracy: less than or equal to
0.25%
1 MΩ / 50Ω selectable load
Sine wave greater than or equal to 200
MHz
BNC (f) to Dual Banana (m) Adapter
50Ω precision feed through
Recommended Model Part
Number
Keysight MSO9254A
Fluke 9500B High Performance
Oscilloscope Calibrator or
Keysight E8257D+1EU option
Keysight 1251-2277
Keysight 0960-0301
Keysight N2818/9A Differential Probes User Guide
Functions
Display probe output
Signal source for DC gain
and bandwidth
Interconnection between
probe and generator
Termination between probe
and calibrator for
bandwidth verification