Service and Troubleshooting
March 2017
6.2.12
Relay output configuration
To change the Normally Open/Normally Closed settings of the K1 and K2 relays, do the following:
1. Disconnect the power supply.
2. Disconnect the relays.
3. Remove the front cover.
4. Remove the plastic display cover.
5. Switch the jumpers to the desired settings: Normally Open (NO) or Normally Closed (NC).
6. Replace the plastic display cover and the front cover.
Note
Ensure that o-rings and seats are in good condition prior to mounting the cover in order to maintain the
specified level of ingress protection.
92
Front cover
Relay K1: NO / NC
Relay K2: NC / NO
Reference Manual
00809-0100-2410, Rev DA
Service and Troubleshooting