Reference Manual
00809-0100-2410, Rev DA
B.6 International
E7
IECEx Flame-proof
Certificate: IECEx FMG 10.0005X
Standards: IEC 60079-0:2007-10,
IEC 60079-1:2007-04,
IEC 60079-7:2006-07,
IEC 60079-11:2006,
IEC 60079-26:2006;
IEC 60079-27:2008
Ex marking
Comm. Bus
Ex d e [ib] IIB T4 Gb
FISCO
Ex d e [ia IIC Ga]
HART/ 4-20mA
IIB T4 Gb
Active
HART/ 4-20mA
Ex d e ib IIB T4 Gb
Passive
Ex d e [ib] IIB T4 Gb
Fieldbus
Special Conditions for Safe Use (X):
1. The flamepaths of the equipment are not intended to be
repaired. Consult the manufacturer if repair of the flamepath
joints is necessary.
Product Certifications
Ui
Ii
Pi
Ci
μ
V
mA
W
F
-
-
-
-
-
-
-
-
30
300
1
0
-
-
-
-
Markings: Ex d e [ib] IIB T4 Gb
Ex d e [ia IIC Ga] IIB T4 Gb
Ex d e ib IIB T4 Gb
Li
Uo
Io
Po
mH
V
mA
W
-
15
354
5.32
-
23.1
95.3
0.55
0
-
-
-
-
15
200
3
Product Certifications
March 2017
Co
Lo
Group
μ
F
mH
-
-
IIB
0.14
3.9
IIC
1.0
15
IIB
3.67
33
IIA
-
-
IIB
1.99
143
IIB
123