Installing the Detector
Flow Connections to the Detector
Preparations
Parts required
When working with solvents please observe appropriate safety procedures (for
WA R N I N G
example, goggles, safety gloves and protective clothing) as described in the
material handling and safety data sheet supplied by the solvent vendor, especially
when toxic or hazardous solvents are used.
The flow cell is shipped with a filling of isopropanol (also recommended when the
N O T E
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
The heat exchanger/capillary and the cell body can be fixed mirror symmetrically to have
N O T E
both capillaries routed to the bottom or to the top (depending on the routing of the
capillaries to the column). For details see
Cell"
38
Detector is installed in the LC system.
Other modules
Parts from accessory kit,
Two wrenches 1/4 – 5/16 inch for capillary connections
on page 114.
see"Detector Accessory Kit Contents"
"Replacing Capillaries on a Standard Flow
1200 Series DAD and MWD User Manual
on page 31.