Installing the Module

Flow Connections to the Detector

Flow Connections to the Detector

For bio-inert modules use bio-inert parts only!
Parts required
#
1
1
1
Sample degradation and contamination of the instrument
C A U T I O N
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
This procedure shows the detector outside of a system. In an Agilent 1260 Infinity Liquid
N O T E
Chromatograph, the detector is located below a G1316 TCC on the bench. In an Agilent 1290
Infinity Liquid Chromatograph, the detector is located between a G1316 TCC (below) and the
Solvent Compartment (above). (See
1
Remove the front cover.
44
Description
System
Max-Light cartridge flow cell
Capillaries and tubing from Accessory Kit.
For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
Do not mix bio-inert and non-inert modules or parts in a bio-inert system.
"Optimizing the Stack Configuration"
2
Remove the black hoods from the cell interfaces (in/out)
and store them in the plastic case provided with the
Max-Light Cartridge Flow Cell.
Agilent 1200 Infinity Series DAD User Manual
on page 32.