Flow Connections to the Detector

Preparations
Parts required
When working with solvents please observe appropriate safety procedures (for
WA R N I N G
example, goggles, safety gloves and protective clothing) as described in the
material handling and safety data sheet supplied by the solvent vendor, especially
when toxic or hazardous solvents are used.
The flow cell is shipped with a filling of isopropanol (also recommended when the
N O T E
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
1200 Series FLD User Manual
Detector is installed in the LC system.
Other modules
Parts from accessory kit,
see"Detector Accessory Kit Contents"
Two wrenches 1/4 – 5/16 inch for capillary connections
Installing the Detector
on page 43.