Operation
Power-Up Standby Screen
Following a successful power-up, the Standby
screen appears and provides instructions on start-
ing the operation of the Fabius OS.
Setting Fresh Gas Flow
Set the fresh gas flow to the desired concentration
using the flow control knobs on the front of the
machine.
S-ORC (Sensitive Oxygen Ratio Controller)
S-ORC is a control element which provides a mini-
mum O
concentration in the fresh gas flow. From
2
a flow rate of approx. 300 mL/min, the N
tration in the fresh gas can be set between 0 and
75 %.
During O
shortage S-ORC limits the N
2
tration in the fresh gas, so that the O
does not drop below 23 vol.%.
S-ORC prevents N
O flow if the N
2
valve is open and O
metering valve closed or if the
2
O
flow is less than 0.2 L/min.
2
70
O concen-
2
O concen-
2
concentration
2
O metering
2
Instructions for Use Fabius OS SW 3.n