Operation

Standby page after start-up

Standby

Setting the fresh-gas flow

The fresh-gas settings can be changed before a
ventilation mode is selected.
 Set the fresh-gas flow.
S-ORC (Sensitive Oxygen Ratio
Controller)
Fabius is equipped with a mechanical minimum O
delivery (S-ORC). This safety device prevents
hypoxic gas mixtures if N
carrier gas.
Starting at a flow of approx. 200 mL/min, the N
concentration in the fresh gas can be set to a value
between 0% and 75%.
When a lack of O
is present, the S-ORC limits the
2
N
O concentration in the fresh gas so that the O
2
concentration will not fall below 23 Vol%.
90
A
B
O is selected as the
2
2
In Standby mode, different instructions (A) and
information (B) are displayed on the screen.
A To start operation press one of the keys
located to the left of the display
B Last system test, last leakage/compliance test,
system leakage, ventilator leakage,
compliance
S-ORC inhibits the N
conditions:
– N
O flow control valve is open, even though
2
the O
flow control valve is closed.
2
– O
flow is below 0.2 L/min.
2
With an N
O failure, O
2
supplied. An alarm is not triggered. The float in the
N
O flow tube drops to zero.
2
2
S-ORC has no oxygen-specific monitoring function
and offers no protection for the consequences of
an accidental interchange of gases.
O
Therefore, the O
monitored.
2
Instructions for use Fabius plus SW 3.n
O flow under the following
2
can continue to be
2
concentration must always be