D
MENU OPTION 3 SETUP MARKER
Two object marker adjustments are available in this menu: Marker Setup, which adjusts the proxim-
ity of the pen tip to the slide for marking, and Marker Calibration, which calibrates the marker.
Marker Setup
Marker Setup adjusts the pen tip nearer to or further from the microscope slide when the pen is low-
ered for marking. Indications that adjustment is needed include:
The pen mark is too wide (too close to the slide)
The pen drags on the slide between marks (too close to the slide)
Scant or no mark on the slide (too far from the slide)
Before changing the Marker Setup, confirm the pen is seated correctly in the holder. Do a Marker Cal-
ibration.
From the menu, select 1, Marker Setup.
A prompt to load a blank slide appears.
Load a blank ThinPrep slide into the slide holder on the stage. Press the Next button on the pod.