2. While maintaining full deflection, roll your finger in a circle 2 or 3 times to move the joystick in
the widest possible circles.
3. Return the joystick to the center and let go.
4. Do not touch the joystick while calibration calculations are being performed.
Joystick
In a few seconds, the calibration is complete. An audible beep will sound and the display indicates
the calibration is complete.
Press the Continue soft key on the key pad to return to the Load Slide screen.
If calibration fails, retry it several times. If calibration continues to fail, contact Hologic Technical
Support.
O
PERATION OF THE
Figure 4-2 Calibrate the Joystick
Joystick Cal.
Complete
Continue
Joystick Cal.
Failed
Continue
S
M
OFTWARE
ENU
Deflect and circle
2 or 3 times
Review Scope Operator's Manual
4
4.5