Hologic ThinPrep Operator's Manual page 70

Hide thumbs Also See for ThinPrep:
Table of Contents

Advertisement

Illuminated
Mark Indicator
Good Alignment
Figure 3-3 Align the Fiducial Mark and the Mark Indicator
When the fiducial mark is aligned with the mark indicator, press the Next button on the pod. The
stage presents the second mark and the display prompts the user to align the second fiducial mark.
Using the joystick control, align the second fiducial mark with the illuminated mark indicator, as nec-
essary. (Pressing the Prev button will allow re-alignment of the first fiducial mark, in case a mistake
was made.) Press the Next button on the pod when satisfied with the alignment.
The third fiducial mark is presented and the display prompts the user to align the third mark.
Using the joystick control, align the third fiducial mark with the illuminated mark indicator, as neces-
sary. Press the Next button on the pod.
O
PERATION OF THE
AUTO REVIEW (continued)
Slide Fiducial Mark
Poor Alignment:
Fiducial Mark is on
top of Illuminated
Mark Indicator
Align 2nd Mark
Press Next/Prev
Align 3rd Mark
Press Next/Prev
R
S
EVIEW
COPE
Poor Alignment:
Fiducial Mark is too far
away from Illuminated
Mark Indicator
Review Scope Operator's Manual
3
3.9