Hologic ThinPrep Operator's Manual page 63

Hide thumbs Also See for ThinPrep:
Table of Contents

Advertisement

3
O
PERATION OF THE
Manual Review
Manual Review refers to a slide review in which
patient slide data is not retrieved from or communicated to the server
a scan of the entire cell spot is conducted by the operator initiating each movement of the
stage
marks may be made (up to 30), but no record is stored on the server
(Refer to Figure 3-1 for a graphical representation of the typical slide review process.)
Important Operational Notes
CAUTION: Never lift or handle the Review Scope by the stage. (Refer to Figure 2-2‚
Moving the Review Scope.)
Calibrate the 4-button pod Joystick and the Object Marker once a day, in that order. (Refer to page 4.4
and page 4.6, respectively.)
Calibrate the 5-button pod joystick upon installation of the pod, or if the pod has been removed and
then reinstalled on a Review Scope. (Refer to page 4.4.)
Load the slide properly, seated fully against the back and left guides, to ensure a successful slide ID
scan and accurate positioning of Object of Interest coordinates sent from the Server. (Refer to Figure
3-2.)
Always examine the entire Field of View. Mark if you see any suspect object and then do Autoscan.
3.2