C h a p t e r T h r e e
Operation of the Review Scope
SECTION
A
OVERVIEW
®
ThinPrep
Pap Test cervical cytology sample slides processed by the ThinPrep Imaging System are
reviewed at the Review Scope (RS). Slide review by a cytotechnologist is assisted by automated fea-
tures. (Refer to Figure 3-1 for typical slide review process.)
Auto Review
In this manual, Auto Review refers to a slide review in which the RS
scans the ID number from the slide
communicates with the server for appropriate slide data record
makes use of the Autolocate function (where the 22 fields of view identified by the Image Pro-
cessor are presented to the cytotechnologist, CT)
makes use of the Autoscan function, as required or desired
uploads slide data record to the server at conclusion of slide review
Subsequent Review
A slide that has undergone Auto Review may be reviewed again, making use of the Autolocate,
review and Autoscan functions. Further marks may be added (to a maximum of 30 marks on a slide),
but no previous marks may be removed. The slide data record will be revised at the Server at the con-
clusion of the review.
Note:
Slides previously screened either via auto review or manually may always be examined again
in manual mode.
O
PERATION OF THE
Review Scope Operator's Manual
R
S
EVIEW
COPE