E Series Operator's Guide
and pacing current are also recorded. If async pace is active, the annotation "ASYNC PACE" is recorded. AED units
additionally include shock count and AED mode annotations.
Actual
event
date/time

Analyze Format

The summary report function records six seconds of pre analysis ECG and 12 seconds of ECG recorded during the
ECG analysis interval with the annotation "SHOCK ADVISED" or "NO SHOCK ADVISED." AED units additionally
include shock count and AED mode annotations. The date/time printed on the top of the strip corresponds to the ECG
data occurring 6 seconds before the event.
Actual
event
date/time
The following annotations may also appear at the top of the Analyze Format printout:
Annotation
POOR PAD CONTACT
ANALYSIS HALTED
NOISY ECG
SHOCK ADVISED
NO SHOCK ADVISED
ECG TOO LARGE
2-8
Description
The MFE Pads are detected as having poor connection.
The ECG analysis is halted due to either the ANALYZE button being pushed or a fault
condition.
Excessive noise is detected.
Shockable rhythm has been detected at the end of user-initiated ECG analysis.
No shockable rhythm has been detected at the end of user-initiated ECG analysis.
The amplitude of the ECG signal is too large for proper rhythm analysis.
9650-1210-01 Rev. R