Temperature Test - ZOLL M Series Service Manual

Hide thumbs Also See for M Series:
Table of Contents

Advertisement

M Series Service Manual

19.0 Temperature Test

Tools Needed: IBP Temp Simulator 9100-0402-TF (or equivalent
Test Setup: None
Do this...
19.1
Connect the 2-channel Y adapter to the
temperature connector (T1/T2) on the M Series
CCT Unit.
19.2
Connect the Temperature Probe Simulator
cables to the 2-channel Y adapter.
19.3
On the Temperature Simulator, set the T1
temperature channel to 59.2 degrees F(15
degrees C).
19.4
On the Temperature Simulator, set the T1
temperature channel to 112.8 degrees F (45
degrees C).
19.5
On the Temperature Simulator, set the T2
temperature channel to 59.2 degrees F(15
degrees C).
19.6
On the Temperature Simulator, set the T2
temperature channel to 112.8 degrees F (45
degrees C).
Record your results on the Maintenance Tests Checklist.
1.
If you are using a Temperature Simulator that does have the same setting above, use the closest value available. Verify the display temperature value
is within 0.2 F or 0.1 C of the simulator setting
1-36
1
)
Verify that...
Verify that the value displayed for the T1 temperature channel between 59.0
degrees to 59.4 degrees F (14.9 to 15.1 degrees C).
Verify that the value displayed for the T1 temperature channel between 112.6
degrees to 113 degrees F (44.9 to 45.1 degrees C).
Verify that the value displayed for the T2 temperature channel between 59.0
degrees to 59.4 degrees F(14.9 to 15.1 degrees C).
Verify that the value displayed for the T2 temperature channel between 112.6
degrees to 113 degrees F (44.9 to 45.1 degrees C).
Pass/Fail
o o o
o o o
o o o
o o o
9650-0450-01 Rev. U