Connecting the Union Tee to the Divert/Inject Valve

Thermo Scientific
To connect the union Tee to the divert/Inject valve
1. Using a fingertight fitting and a ferrule, connect a length of red PEEK tubing to port 3 of
the divert/inject valve. Or, use a stainless steel nut and ferrule to connect the tubing to the
divert/inject valve.
2. Using a fingertight fitting and a ferrule, connect the other end of the tubing to the free
end of the union Tee. See
Figure 29. Six-port divert/inject valve connections
To
LC TEE union
4
5
Plug
(optional)
Detector position
Figure 29
and
Figure
30.
To
union Tee
From
3
LC
2
1
Plug
(optional)
6
To
waste
6
Connecting the Inlet Plumbing
Setting Up the Inlet for High-Flow Infusion
From
3
LC
4
2
5
1
6
To
waste
Waste position
LTQ Series Getting Connected Guide
47