Connecting the MS Detector to the Data System Computer

Thermo Scientific
The data system for the LTQ Series MS detector consists of a computer, a monitor, and an
optional printer. The LTQ Series MS detector communicates with the data system computer
through an Ethernet cable.
To connect the MS detector to the data system computer
1. Connect a category five network (Ethernet) cable (P/N 00302-01838) to the
ETHERNET 100 BASE-T connector located on the MS detector power panel.
2. Connect the opposite end of the Ethernet cable to the 10/100 Base-T Ethernet switch
(P/N 00825-01015) provided with the MS detector.
3. Connect a second Ethernet cable (P/N 00302-01838) from the Ethernet switch to the
Ethernet card on the data system computer labeled Surveyor MS.
4. Connect the Ethernet switch to line power.
1
Line Power, Vacuum System, Gases, and Ethernet Communication
Connecting the MS Detector to the Data System Computer