4
Connecting External Devices
External Devices Not Controlled by the Xcalibur Data System
Figure 16. Schematic of the contact closure connection between the MS detector and an external device
Peripheral Control
Start
Ready
In
Peripheral Control
Start
Ready
In
Analog Input

Starting a Sequence Run from Xcalibur

28
LTQ Series Getting Connected Guide
Start
Out
Out
Start
Out
Out
1V Max
10V MAX
When the Xcalibur data system does not control the autosampler you are using, Xcalibur
selects the LTQ Series MS detector as the start instrument for a sequence run. When you are
ready to inject a set of samples, ensure that the LTQ Series MS detector is not listed as the
start instrument in Xcalibur.
The LTQ Series family of MS detectors consists of the LCQ Fleet, LXQ, LTQ, and LTQ XL
MS detectors.
To start the sequence run
1. On the Xcalibur Roadmap – Home Page, click the Sequence Setup button to open the
Sequence Setup window.
2. Open the sequence that you want to run:
a. Choose File > Open.
The Open dialog box appears.
b. Browse to the appropriate folder, and then select a sequence file.
c. Click Open to open the sequence and exit the Open dialog box.
Sequence files are identified by their .sld file extension.
3. Choose Actions > Run Sequence or Actions > Run This Sample.
The Run Sequence dialog box shown in
Power In
230 V AC
10A
Figure 17
appears.
Inlet device
Contact closure
relay contact
TTL logic
Thermo Scientific