5
Connecting a Thermo Scientific LC System
Making Contact Closure with a Surveyor Plus LC System
Figure 23. Surveyor Plus LC system connected to an LTQ Series MS detector
MS Detector
Autosampler
7-connector LC/MS interconnect cable
Start In
+ -
Adapter cable for Surveyor MS Pump
7 in. (18 cm) P/N 60053-63038
To
LC/MS interconnect
cable connector
labeled PUMP
Not drawn to scale
36
LTQ Series Getting Connected Guide
3. To connect a Surveyor MS Pump or Surveyor MS Pump Plus to the LC/MS interconnect
cable:
a. Connect the end of the adapter cable labeled PUMP to one of the PUMP connectors
of the 7-connector LC/MS interconnect cable.
b. Plug the end of the adapter cable labeled MS PUMP into the 8-pin socket on the rear
panel of the Surveyor MS Pump or Surveyor MS Pump Plus.
Detector
A/S
Pump
LTQ Series MS detector
(right-side panel)
To
Surveyor MS Pump
Detector
Detector
Pump
LTQ/LXQ
interconnect
adapter cable
Solvent platform
Detector
(rear panel)
Autosampler
(rear panel)
Surveyor
LC Pump
(rear panel)
Connector labeled PUMP
OR
Connector labeled PUMP
Surveyor
MS Pump
Connector
(rear panel)
labeled
MS PUMP