Line Power, Vacuum System, Gases, and
Ethernet Communication
Connecting the MS Detector to Line Power
Thermo Scientific
This chapter describes how to connect the LTQ Series MS detector to line power, the
forepump, necessary gases, and the data system computer.
Contents
•
Connecting the MS Detector to Line Power
•
Connecting the Forepump to the LCQ Fleet or LXQ MS Detector
•
Connecting the Forepumps to the LTQ XL MS Detector
•
Connecting the Gases to the MS Detector
•
Connecting the MS Detector to the Data System Computer
To connect the LTQ Series MS detector to line power
1. Turn the Main Power circuit breaker to the Off (O) position.
2. Make sure that the Electronics switch is in the Service Mode position.
3. Connect the female plug of the power cord to the POWER IN receptacle of the power
panel.
4. Connect the male plug of the power cord to the 230 V ac power source in your
laboratory.
CAUTION If your local area is subject to power fluctuations or power interruptions, you
must install a power conditioning device or an uninterruptible power supply (UPS) in
your laboratory. For more information, refer to the LTQ Series Preinstallation Requirements
Guide. The UPS must be certified by both North American (UL, CSA) and European
Agencies (TUV, UDE, SEMKO, DEMKO, and so on).
1
LTQ Series Getting Connected Guide