Connecting the Inlet Plumbing
Setting Up the Inlet for High-Flow Infusion
Figure 30. Connecting the union Tee to the divert/inject valve
3
4
5
6

Connecting the LC Pump to the Divert/Inject Valve

Connecting the Divert/Inject Valve to a Waste Container

48
LTQ Series Getting Connected Guide
Stainless steel nut
(P/N 2522-0066)
Stainless steel ferrule
(P/N 2522-3830)
2
1
PEEK tubing
(P/N 00301-22912)
LC TEE union
To connect the LC pump to the divert/inject valve
1. Using a fingertight fitting and a ferrule, connect a length of PEEK tubing to port 2 of the
divert/inject valve.
2. Using an appropriate fitting and ferrule, connect the other end of the tubing to the outlet
of the LC.
To connect the divert/ inject valve to a waste container
1. Using a fingertight fitting and a ferrule, connect a length of red PEEK tubing to port 1 of
the divert/inject valve.
2. Insert the other end of the tubing into a suitable waste container.
Infusion line
LC union
Figure 29
shows the ports of the divert/inject valve.
Figure 29
shows the ports of the divert/inject valve.
Syringe