Flow-Mode Sampling - Teledyne 3700 Installation And Operation Manual

Portable samplers
Table of Contents

Advertisement

3700 Portable Samplers
Section 3 Programming Guidelines
Example: Save Current Program (Continued)
4. You can save the current program only when the sampler is configured for the extended programming mode. To
verify the programming mode setting, press the
5. Select "EXTENDED." Press the
6. Press the
RIGHT ARROW
/
7. Press the
ENTER
PROGRAM
8. Select the number you want to use as the program "name." Select "NONE" when you do not want to save a
program When the sampler has saved the program, it will display the next configure option. When you are done
configuring the sampler, press
9. When the sampler is properly installed, press the

3.9.9 Flow-Mode Sampling

3-46
SELECT OPTION: (← →)
SET CLOCK
ENTER
SELECT OPTION: (← →)
PROGRAMMING MODE
/
key to accept the selection.
ENTER
PROGRAM
[BASIC,
PROGRAMMING MODE
key to skip the Load Stored Program configure option.
SELECT OPTION: (← →)
LOAD STORED PROGRAM
key to access the Save Current Program configure option.
SELECT OPTION: (← →)
SAVE CURRENT PROGRAM
/
to return to Standby.
EXIT
PROGRAM
SAVE PROGRAM AS
[ #1 , #2 , #3 , NONE ]
START SAMPLING
. . . STANDBY . . .
10:37:23
The Flow-Mode Sampling option directs the sampler to take a
sample at key times in a flow-paced sampling program. There
are two settings. The first setting directs the sampler to take a
sample at the start time. The second directs the sampler to take
a sample at the switch times (when the bottle set changes are
based on time). Both settings of the option are disregarded in
STORM sampling routines. (Display #40)
Sample at Start Time – If you select "YES," the 3700 will take
the first sample at the start time entered in the program
sequence. If you select "NO," the first sample is delayed until the
number of flow pulses, set in the program sequence, has counted
down to zero. (Display #270)
Sample at Time Switch – If you select "YES," the sampler will
switch to a new bottle or set, take a sample, and reset the flow
pulse count. The next sample will occur after a full pulse
countdown. If you select "NO," no samples be taken at the switch
time and the pulse countdown will not be reset. The switch will
/
key.
PROGRAM
EXTENDED
]
key to run the program.
19-JUN-04