Teledyne 3700 Installation And Operation Manual page 71

Portable samplers
Table of Contents

Advertisement

1. Press
/
ENTER
PROGRAM
2. Access the program sequence by selecting "PROGRAM."
3. Step through the program until the "CALIBRATE SAMPLER?" input display appears (Step 4).
Other program sequence displays.
4. Select "YES."
5. Before pressing the
MANUAL SAMPLE
The sampler will deliver the programmed sample volume.
6. Measure the actual volume delivered and enter that value here.
7. Repeat the procedure if desired by selecting "YES." When you are satisfied with the calibration, select "NO."
Under normal conditions, you should not need to repeat the procedure.
each pumping cycle. Then, check the suction-line entries in the
configure sequence to see that they are accurate. The calibration
procedure is intended to be for "fine tuning" only.
After you enter the sample volume actually delivered, as shown
in Step 6, all subsequent sample volumes delivered will be
adjusted to correct for the difference between the expected
sample volume and the actual volume entered. Clear the
adjustment by changing a suction line entry in the Suction Line
configure option or by reinitialization of the sampler.
Because the sample volume can be calibrated to ± 10 ml, use a
graduated cylinder to facilitate measurement. Teledyne Isco
offers a graduated cylinder and spare bottles as accessories from
the factory. Refer to the Appendix B Accessories List for details.
The calibration pump cycle will include rinse cycles and retries, if
the sampler is configured to perform them. This ensures that the
calibration procedure includes the pump cycle used while
running the sampling program.
Example: Calibration
to access the interactive state.
. . . STANDBY . . .
9:34:50
[
PROGRAM
SAMPLER
[
TIME
PACED SAMPLING
CALIBRATE SAMPLER?
[
YES
key, make sure a collection container is underneath the distributor.
PRESS MANUAL SAMPLE
KEY WHEN READY . . .
. . . MANUAL SAMPLE . . .
PUMPING
200
ml VOLUME
DELIVERED
CALIBRATE SAMPLER?
[ YES ,
Section 3 Programming Guidelines
19-JUN-04
, CONFIGURE ]
, FLOW]
, NO ]
200 ml
]
NO
3700 Portable Samplers
3-37