Teledyne 3700 Installation And Operation Manual page 49

Portable samplers
Table of Contents

Advertisement

Tests indicate that the pump produces sound levels in excess
of 85db at one meter. Prolonged exposure to this sound level
could result in hearing loss and requires the use of protective
ear plugs.
• When you press the
message similar to the one shown below. This message
indicates the total number of pump counts that occurred
while the pump was running. This information can be
used to set the desired number of purge counts required
to purge the suction line. See the Purge Counts configure
option. Press any key, except
to the standby display..
311 COUNTS FORWARD
- The
key will stop the pump any time it is
STOP
STOP
running. When you press the
the sampling routine will halt, the sampler will transfer
to the standby state, and the "PROGRAM HALTED"
message will be displayed. Pressing the
in the interactive state will access the display's
reference number.
START SAMPLING
pressing the
START SAMPLING
sampling program. When entering a sampler ID, this
key will type a space.
RESUME SAMPLING
displayed, pressing the
the sampler to continue with the current sample
program at the point were it was halted. When entering
a sampler ID, this key will type a period.
MANUAL SAMPLE
will allow you to take a manual sample. The
key is valid in the standby state, the run state,
SAMPLE
and when calibrating the sampler. If you press the
key while a sampling program is in
MANUAL SAMPLE
progress, the manual sample will be counted as one of
the programmed samples.
When entering a sampler ID, this key will type a dash
(-).
- If you want to reposition the
NEXT BOTTLE
distributor, use the
advance the distributor to the next bottle position. (The
sampler determines the bottle positions by referring to
the Bottle Number configuration setting.) The sampler
will display the distributor's location as illustrated in g.
After a short time, the display will revert to standby. If
3700 Portable Samplers
Section 3 Programming Guidelines
CAUTION
key, the sampler will display a
STOP
and
/
STOP
ON
OFF
key in the run state,
STOP
STOP
- When in the standby state,
key will begin the
- When "PROGRAM HALTED" is
key will cause
RESUME SAMPLING
- Pressing the
MANUAL SAMPLE
key. The sampler will
NEXT BOTTLE
, to return
key while
key
MANUAL
3-15