Power – The sample event was initiated to compensate for a
missed sample which should have occurred while the sampler
was without power. If more than one sample was missed, only
one sample will be taken.
Enable – The sample event was initiated when the sampler
became enabled by a device connected to pin F of the flow meter
connector, generally a 4200 Series Flow Meter, a 4100 Series
Flow Logger, or Liquid Level Actuator. This source is also
reported when the sample was initiated at a programmed
resume time. Two different configure option settings control this
event: the Sample Upon Enable setting in the Enable Pin con-
figure option and the Sample at Resume setting in the Sampling
Stop/Resume configure option.
Manual – The sample event was initiated with the manual
sample key and was counted as one of the programmed sample
events.
Time Switch – The sample event was initiated at the pro-
grammed switch time. This event is controlled by the Sample at
Time Switch setting in the Flow Mode Sampling configure
option. This option is available for flow-paced sampling only.
Disable – The sample event was initiated when the sampler
became disabled by a device connected to pin F of the flow meter
connector, generally a 4200 Series Flow Meter, a 4100 Series
Flow Logger, or a Liquid Level Actuator. This source is also
reported when the sample was taken at a programmed stop time.
Two different configure option settings control this event: the
Sample Upon Disable setting in the Enable Pin configure option
and the Sample at Stop setting in the Sampling Stop/Resume
configure option.
3.17.5 Error Messages and
The probable cause of any missed sample is reported after the
Missed Samples
sample number/source display. A missed sample display is illus-
trated below.
SAMPLE MISSED:
NO LIQUID DETECTED!
Twelve causes are tracked and reported:
Pump '
' Key Hit! – The sampler was halted with the stop
STOP
key during the sample event.
Pump Jammed! – The sampler was unable to take the sample
because the pump jammed.
Started Too Late! – The sampling routine was started after the
programmed start time for the first sample. This message is
reported for all samples skipped because of an expired start time.
Program Halted! – The sampling routine was interrupted by
the
or
key when the sample event was to occur.
STOP
EXIT PROGRAM
Power Lost! – The sampler was unable to take the sample
because the sampler's power source was disconnected.