Teledyne 3700 Installation And Operation Manual page 68

Portable samplers
Table of Contents

Advertisement

3700 Portable Samplers
Section 3 Programming Guidelines
Example: Time-Paced Composite Sampling
1. If the sampler is not already on, press the
interactive state.
2. Access the configure sequence by selecting "CONFIGURE."
3. Press the
LEFT ARROW
configure option appears.
/
4. Press the
ENTER
PROGRAM
5. Select "PORTABLE" for either the 3700 Standard or Compact.
6. Select "1" for single bottle composite sampling.
7. Enter the bottle size here, "15,000." (Table 3-2 lists the standard Teledyne Isco bottle sizes.)
8. Press the
/
ENTER
PROGRAM
3-34
Example Time-Paced Composite Sampling demonstrates the pro-
gramming procedures for a composite sampling routine. When
composite sampling with the 3700 Standard, you must convert
the unit for single-bottle sampling. With the 3700 Compact, you
must install the proper bottle into the base.
After converting the 3700 Standard, check the settings to verify
the sampler is configured for composite sampling. This example
begins with a demonstration of the steps needed to check the
Bottles and Sizes configure option settings. The Bottles and Sizes
configure settings should indicate that only one bottle is being
used. Be sure to enter the correct bottle volume. The settings
used in this example are for the 4-gallon (15,000 ml) composite
bottle available with the 3700 Standard.
This time-paced routine will take samples at 15-minute intervals
for a 12-hour period. Sample volumes are to be 250 ml each.
When entering the program settings, you must enter the number
of samples required before entering the sample volumes. At four
samples per hour, the 12-hour period would yield 48 samples.
The total sample volume collected at the end of the routine would
be 12,000 ml (48 samples
capacity of the 4-gallon bottle.
/
ON
OFF
. . . STANDBY . . .
10:34:50
[ PROGRAM,
SAMPLER
or
key to scroll through the configure options until the Bottles and Sizes
RIGHT ARROW
SELECT OPTION: (← →)
SET CLOCK
key to access the Bottles and Sizes settings.
SELECT OPTION: (← →)
BOTTLES AND SIZES
[
PORTABLE
SAMPLER
[
4, 12, 24 ]
1
,
BOTTLES
BOTTLE VOLUME IS
15000
key at this display to access the Suction Line input displays.
250 ml/sample), well within the
x
key to turn it on. Press
ENTER
19-JUN-04
]
CONFIGURE
, REFRIG
ml
/
to access the
PROGRAM