1. Press
/
ENTER
PROGRAM
2. Access the program sequence by selecting "PROGRAM."
3. Select "STORM.
After displaying this message briefly, the sampler will automatically advance to the next display.
4. Enter "25" to delay the first group sample 25 minutes. (Acceptable range of values: 0 to 999.)
5. Enter "2" to take two timed samples.
6. Enter "5" to set the time interval between samples to five minutes.
7. Enter the number of bottles per sample event: "3."
8. Enter "1" to place one sample in each bottle.
9. Enter the sample volume of "900" ml.
The sampler will display this message before automatically advancing to the next display.
10. Select "FLOW."
11. Select "DURING" to fill the second bottle group at the same time as the first bottle group.
12. Enter "5" as the flow-pulse interval. The sampler will take a sample every fifth flow pulse.
13. This display requires two entries. Enter "2" hours and "0" to set the maximum amount of time between flow-paced
sample events.
Example: Storm Sampling
to access the interactive state.
. . . STANDBY . . .
10:34:50
[
PROGRAM
SAMPLER
[ TIME , FLOW ,
PACED SAMPLING
TIME MODE
FIRST BOTTLE GROUP
25
MINUTE DELAY TO
FIRST GROUP SAMPLE
TAKE
EVENTS ( 1 - 50 )
SAMPLE INTERVALS OF
5 MINUTES (1 - 999)
3
BOTTLES PER
SAMPLE EVENT ( 1-23 )
1
SAMPLES PER
BOTTLE (1 -
SAMPLE VOLUMES OF
900
ml
18 BOTTLES, 18000 ml
AVAILABLE
[TIME,
SECOND BOTTLE GROUP
[
DURING
TIME MODE
SAMPLE EVERY
PULSES ( 1-9999 )
5
Section 3 Programming Guidelines
12-JUN-04
, CONFIGURE ]
STORM
]
2
TIMED
2)
( 10 - 1000 )
FLOW
]
, AFTER]
3700 Portable Samplers
Display 1
Display 12
Display 15
Display 23
Display 24
Display 30
Display 31
Display 50
Display 14
Display 16
Display 22
3-95