Storm Sampling: Time Pacing For The Second Bottle Group - Teledyne 3700 Installation And Operation Manual

Portable samplers
Table of Contents

Advertisement

3700 Portable Samplers
Section 3 Programming Guidelines
Figure 3-17 Storm Sampling: Time Pacing for the Second Bottle Group
The "TIME MODE" message marks the beginning of the first
section. This message is followed by Display #15, below.
Display #15
-- MINUTE DELAY TO
FIRST GROUP SAMPLE
Use this display to set amount of time between the time the
sampler is enabled and the first sample delivered to the first
bottle group. (The sampler may be enabled at the start time.)
The acceptable range of values for the delay to first group sample
is 0 to 999 minutes. If the delay is 0, the sampler will take the
first sample when enabled by the flow meter. Subsequent
time-paced samples will occur at the time interval entered in
Display #23.
3-73

Advertisement

Table of Contents
loading

Table of Contents