4.1 Daily User Inspection and Maintenance

During maintenance and inspection, strictly observe precautions contained in "Chapter 1 For Safe
Operation" in this manual for you to use the DR-ID1270 under best conditions.

4.1.1 Periodical Inspection

Inspection Every Three Months
Once every three months, remove any dirt or dust accumulated in each part of the equipment using
a vacuum cleaner or air duster, clean each part with a slightly moistened soft cloth and then wipe off
any moisture with a dry cloth.
See "2.2 Unit Names and the Functions" (page 2-3).
CAUTIONS
Be sure to turn off the power before cleaning each part of the device.
No.
Unit
1
Flat panel sensor
No.
Unit
4
Power box
CAUTIONS
Ensure sufficient space when cleaning the equipment on a table, etc.
Air filter
Clean the air filter on the rear of the power
supply unit with a vacuum cleaner.
Remove the louver while pressing the lever
on the right side, and then clean the air filter
with a vacuum cleaner after detaching it from
the louver.
No.
Unit
2
Power supply unit
No.
Unit
3
Power supply unit
Air filter (1)
Air filter