Startup

3.5 Home screen data views

3.5.1 Current tab

Page 46
Diagnostics icon allows you to check the health of the analyzer. Touching the icon takes
you to the diagnostics dialog box without interrupting the current analysis.
Data review icon allows you to view the information from the log file. Touching the icon
takes you to the data review dialog box without interrupting the current analysis.
Sign on or sign off icon allows you to sign onto the analyzer to modify or change the
operation, or to sign off to prevent changes, without interrupting the current analysis.
If no user is signed on, the sign on icon is present. Touching the icon when no one is
signed on causes the sign on dialog box to appear, so you may enter your user ID and
password.
If you are already signed on, the sign off icon is present. Touching the icon when you
are signed on signs you off. While you are logged off, you have view-only access to all
dialog boxes except security and the audit trail, but you can issue print commands.
Home screen data views include current, log view, and graph tabs. You may switch from
tab to tab without interrupting the current analysis.
The current tab displays the last analysis values, depending on the selected analysis
mode. In all modes except standby, offline, conductivity, and clean, the current tab
displays the last readings for TOC, conductivity, temperature, curve type, trend, and the
time of the last reading. The home screen also indicates the current state of the analysis
in text and through process animation. See
Text field (left side of screen) includes the following variables:
Analysis state is the current state (idle, flushing, stabilizing, analyzing, or reporting).
Run mode is the selected mode of the analyzer. The choices for operation mode are
online TOC, conductivity, standby, offline, self clean, and online manual TOC.
TOC reading is in "TOC: X.X ppb" format. The range of values is 0.0 to 2000 ppb. If
the value exceeds 2000, "Over limit" is displayed.
Conductivity is the last conductivity reading in "Conductivity: X.XX µS/cm" format for
compensated conductivity and "Conductivity: X.XX µS/cm U" format for
Figure
20.
Figure 20 Home screen, current tab