Dimensions in
[228.6]
9
2.2.1 Mounting general considerations
[mm]
in.
Figure 3 PAT700 analyzer orientation
Follow these guidelines when installing the analyzer:
•
Locate the analyzer where it is accessible for operation, service and calibration.
•
Minimize the distance between the water system sample point and the analyzer to
allow for representative sampling.
•
Maintain sufficient space to allow access to the power switch, and the plumbing and
electrical connections on the end panels.
•
Locate the analyzer where the ambient temperature remains from 10 to 40 °C
(50 to 104 °F).
•
The mounting bracket is installed independently from the analyzer. Once mounted,
the analyzer hooks on the mounting bracket and locks into place with twist latches.
•
Mount the mounting bracket to a stable, flat surface or dual instrument poles.
•
The front of the analyzer requires 12 inches (30.48 cm) of clearance to allow the doors
to open completely.
•
The analyzer is available with ¾-inch conduit openings or quick-connect fittings for
power and I/O wiring. If conduit is used, install fittings that ensure a complete seal with
the openings and are properly sealed to keep the analyzer electronics compartment
watertight.
[597]
23.5
Installation
UP
Page 15