Section 8

8.1 Accessing the diagnostics dialog box

8.2 Diagnostics dialog box

8.3 General tab

Diagnostics
At the home screen, touch the << icon to open the sliding toolbar, then touch the
diagnostics icon, or
Touch the dual UV lamp status indicator in the lower right corner of the screen.
The diagnostics dialog box provides access to lamp management, system tests, and
calibration dates.
The general tab displays the operating status of lamps 1 and 2. See
Figure 101 Diagnostics dialog box, general tab
Version information text label displays the firmware and operating system version
information.
Status icon represents the status of each of lamp 1 and lamp 2. If a lamp is not currently
in use, the icon is dimmed.
Good (green) means the lamp has not reached 0 hours left and is producing a
sufficient level of UV light based on the UV Detect™ diagnostics.
Marginal (yellow) means the lamp has either reached 0 hours left or has been
determined by UV Detect diagnostics to be producing a diminished level of UV light
and that the lamp should be replaced.
Failed (red) means the lamp has failed and is not longer producing sufficient UV light
to perform full sample oxidation.
Not in use (dimmed) means the lamp is not in use.
Hours left text label counts down the number of hours remaining for the UV lamp. You
should replace the UV lamp after 4300 hours of operation.
Figure
101.