User Manual
DOC022.53.80400
Pocket Pro
Salinity
1

Advertisement

Table of Contents
loading

Summary of Contents for Hach 9531600

  • Page 1 User Manual DOC022.53.80400 ™ Pocket Pro Salinity...
  • Page 2: Specifications

    Specifications Specifications are subject to change without notice. Specification Details Dimensions (W x D x H) 37 x 30 x 170 mm (1.45 x 1.18 x 6.69 in.) Enclosure rating IP67 Weight 135 g (0.297 lb) with batteries Power requirements (internal) AAA alkaline batteries (4x) Battery life 450 hours...
  • Page 3: Precautionary Labels

    C A U T I O N Indicates a potentially hazardous situation that may result in minor or moderate injury. N O T I C E Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special emphasis. Precautionary labels Read all labels and tags attached to the instrument.
  • Page 4: Product Components

    Figure 1 Product features 1 Sensor cap 3 Keypad 5 Lanyard attachment 2 Sensor 4 Display 6 Pocket clip Product components Make sure that all components have been received. Refer to Figure 2. If any items are missing or damaged, contact the manufacturer or a sales representative immediately.
  • Page 5 W A R N I N G Fire hazard. Battery substitution is not permitted. Use only alkaline batteries. Install the four AAA alkaline batteries in the tester. Refer to the illustrated steps in Figure Figure 3 Install the batteries User interface and navigation Display description Figure 4 shows the measurements, calibration standard information and indicator icons shown on the display.
  • Page 6: Icon Description

    Table 1 Display icons Icon Description Shows the battery power level. Flashes when the battery power is less than 10%. Shows when the lock feature is on. When the lock feature is on, the parameter value on the display does not change. Push to set the lock feature to on or off.
  • Page 7: Advanced Operation

    Calibrate Items to collect: One auto-recognition calibration standard 1. Set the power to on. 2. Remove the cap from the sensor. 3. Push to go to calibration mode. The auto-recognition standard(s) the tester expects to measure shows on the bottom line. 4.
  • Page 8: Maintenance

    Maintenance C A U T I O N Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document. Clean the sensor Clean the sensor when: • Stabilization is slow • Results drift or are not accurate 1.
  • Page 9 Table 4 Replacement parts Description Quantity Item no. AAA alkaline batteries 4/pkg 4674300 Lanyard 201305 SINGLET, 12.88 mS/cm 20/pkg 2771520...
  • Page 10 Tel. +49 (0) 2 11 52 88-320 SWITZERLAND Fax (970) 669-2932 Fax +49 (0) 2 11 52 88-210 Tel. +41 22 594 6400 orders@hach.com info-de@hach.com Fax +41 22 594 6499 www.hach.com www.de.hach.com © Hach Company/Hach Lange GmbH, 2013–2014, 2023. All rights reserved. 01/2023, Edition 3 *DOC022.53.80400*...