    Index 1. Technical Specifications . . . . . . . . . . . . . . . . . . 2. General information .
  • Page 6: Technical Specifications

    1. Technical Specifications Measuring ranges: Ambient conditions -2 .00 to 19 .99 Working temperature 5 to 40 ºC (41 to 104 ºF) . ± 2000 Storage temperature -15 to 65 ºC (5 to 149 ºF) . Temp -20 to 150 ºC (-4 to 302 ºF) Relative humidity <...
  • Page 7: General Information

    2. General information 2.2. Description As a result of constant improvements to our products sometimes differences may exist between this manual and the instructions supplied with the instrument . The sensION is an instrument to measure pH, mV or ™ PH 3 temperature .
  • Page 8: Installation

    3. Installation DANGER Only qualified personnel, following the local security regulations, should conduct the tasks described in this section of the manual . 3.1. Assemblage 1. Unpack the instrument . 2. Verify that the shift is in 6. Place the electrode on the same position than the electrode’s support .
  • Page 9: Connections

    4. Operation 3. Installation 3.2. Connections 4.1. Description Keys DANGER Make sure the power adapter matches your local power Switch on / Switch off . supply . pH measurement . pH / mV REFERENCE Pt 1000 Pt 1000 STIRRER ELECTRODE ELECTRODE PROBE PROBE...
  • Page 10: Start-Up

    4. Operation 4.3. General diagram 4.2. Start-up Connect to the rear panel of the instrument: This is a general view over the way to access to all possibilities of - the pH electrode the instrument: - the temperature probe (A .T .C .) built-in the pH electrode or - DISPLAy CONTRAST as separated sensor - LANGUAGE...
  • Page 11: Ph And Mv Calibration

    23˚C The calibration frequency depends on the required accuracy and Time 19 s the effect of the samples on the electrode . Hach recommends daily calibration, but users should decide on the right time based on their own experience .
  • Page 12: Ph And Mv Measurements

    4. Operation 4.5. pH and mV measurements 4.6. Other options Measurement by stability - DATE AND TIME The value is fixed on the display when during 6 seconds the - LANGUAGE reading variation is less than 0 .01 pH (0 .5 mV) . - DISPLAy CONTRAST Continuous measurement 10-04-2010...
  • Page 13: Recognized Buffers

    6. Maintenance 4. Operation 4.7. Recognized buffers 6.1. Clean the instrument Buffer solutions Important note: Never use cleaning such as turpentine, acetone The sensION has in memory the following data (pH or similar products to clean the instrument including the display ™...
  • Page 14: Warning Messages

    6. Warning messages 7. Accessories and spares During calibration Code Description LZW9400 .99 Standard redox solution of 220 mV, 250 ml flask . CALIBRATION CALIBRATION MEASURE STABLE 1st buffer 1st buffer LZW9463 .99 pH buffer solution 4 .01, 250 ml flask . ATENTION ! ATENTION ! pH 12.78...
  • Page 15: Limited Warranty

    The remedies of repair, replacement or refund of purchase price as stated above are the exclusive remedies for the breach of this warranty . On the basis of strict liability or under any other legal theory, in no event shall Hach Company be liable for any incidental or...
  • Page 17: Contact Information

    Contact Information HACH Company Repair Service in the Repair Service in Canada: Repair Service in World Headquarters United States: Hach Sales & Service Latin America, the P.O. Box 389 HACH Company Canada Ltd. Caribbean, the Far East, Loveland, Colorado Ames Service...